Quality Control Reports

Internal QC runs, Levey-Jennings charts, and Westgard rule violations

QC Runs Today

7

Across all instruments

Passed / Warnings

5 Pass · 1 Warn

Within acceptable limits

Failures

1

Westgard rule violations — review before reporting

1 QC failure detected. Patient results on the affected analyser must be held until the issue is resolved.

Today's QC Runs

Analyte / LevelInstrumentMeanSDCVResult

Glucose

Level 1 (Low)

Cobas 600095.22.12.2%
Pass

Glucose

Level 2 (Normal)

Cobas 6000188.54.22.2%
Pass

Haemoglobin

Level 1

Sysmex XN-10008.80.151.7%
Pass

Haemoglobin

Level 2

Sysmex XN-100014.20.221.5%
Pass

TSH

Level 1

Architect i20000.450.036.7%
Warning

2S Rule Trigger

Troponin I

Level 1

Architect i20000.0120.001512.5%
Fail

10x Rule Violation

Creatinine

Level 2

Cobas 60005.80.183.1%
Pass

Levey-Jennings Chart